Print ISSN:-2249-8176

Online ISSN:-2348-7682

CODEN : PJMSD7

Current Issue

Year 2024

Volume: 14 , Issue: 3

  • Article highlights
  • Article tables
  • Article images

Article Access statistics

Viewed: 196

Emailed: 0

PDF Downloaded: 400


Kumaran, Srinivasan, Anandh, Shankar, Anbarasan, Muthuselvan, and Suresh Kumar P: Relationship of body mass index on the radiation exposure of patient in diagnostic coronary angiogram through radial route by dose area product


Introduction

Ionising radiation during diagnostic and therapeutic cardiology and radiological procedures are very common in present day medical practice and are one of the largest medical sources of radiation to humans.1 Increasing prevalence of obesity in the general population, obesity has become more prevalent among patients undergoing cardiac catheterisation.2 Increasing body mass Index results in higher patient radiation doses during coronary angiography.2 The greatest source of physician radiation exposure during cardiac catheterisation comes from the patient, which itself is proportional to patient radiation dose.2 The raise in obesity has impacted radiation usage in the catheterisation laboratory because obese patients undergoing fluoroscopic procedures receive greater radiation doses than non obese patients.3, 4, 5 Increase in radiation is attributable to the increased energy required to overcome tissue attenuation and facilitate a sufficient number of photons reaching the image intensifier to generate adequate images.6 The greatest source of physician radiation exposure during cardiac catheterisation comes from scatter radiation. Chronic repeated exposure to low doses of ionizing radiation that is thought to account for increased incidence of premature cataract formation. 7, 8 The dose area product (DAP), related to the effective dose, is a measure of stochastic risk and a potential quality indicator.9 Very simple radiation reduction techniques enabled us to reduce median dose area product for below the values for coronary angiography and percutaneous coronary intervention.10, 9 Procedures that utilize ionizing radiation should be performed in accordance with the As Low As Achievable (ALARA) principle. Fluoroscopy time is a non- dosimetric quantity. It is readily available.

Nevertheless it does not incorporate information about dose rate and skin entrance ports.11, 12 Radial approach when compared to femoral approach, former one carried more radiation dose.13, 14 Novel noise fluoroscopy system, such as those utilize real time image noise reduction technology, have been demonstrated to reduce radiation doses by ≈ 50% or more. 15, 16, 17, 18 Evidence of a detectable acute DNA damage response in operators performing fluoroscopic procedures has been demonstrated. 19 Long term radiation exposure among interventional cardiologists has been linked to multiple adverse effects. 20, 21

Aim

The aim of the study is to find out the effect of body mass index on the radiation exposure of patient in radial coronary angiogram.

Objective

As quoted by previous studies, body mass index has very high influencing factor on the radiation exposure of patient. Our aim is to find out BMI effect on patient radiation exposure exclusively on patients who undergo radial coronary angiogram. Recently as we are getting more obese patients with coronary artery disease as the whole world is going through pandemic of obesity. As such obesity is responsible for procedure related complications from vascular access to final outcome of the procedure. Since radiation exposure has many short term to long term effects, like skin rash to cancer. This study intended to the effect of BMI on radiation exposure. We have chosen patients who undergone coronary angiogram exclusively through radial access as it is the preferred one when compared to femoral access. Illustrative (Figure 1 A and B) is shown as examples from our study.'At the same time radial access coronary angiogram cause more radiation when compared to femoral access.

Study design

Retrospective, single centre observational analytical study.

Materials and Methods

Study was conducted in Thanjavur Medical College in the Dept. of Cardiology. From the coronary angiogram and case history of 77 coronary heart disease admitted in the cardiology department between month of July and August of 2021. Angiogram was done in Single plane catheterisation lab Phillips Allura XD 10. Body mass index was measured using patients height in metres and weight in Kilograms. BMI = Kg/ m2. z Patient radiation exposure was measured by the catheterisation lab system and the radiation metrics like fluoroscopy time, Air kerma and Dose area product are taken from radiation log of Individual procedure for that particular patient. Physician radiation dose as measured by Thermiluminescent dosimeter (TLD) Badge was not taken as it is beyond the scope of this study. Study group was divided into four groups with BMI. a < = 21 b.21.01 to 25.0. c. 25.01 to 29.00. d. > 29.01. Measurement of patients’ radiation exposure was assessed by Dose Area Product which was measured in Gy-cm2. Fluoroscopy time of procedure also measured. Baseline characteristics like age, gender, co-morbidities blood pressure, ejection fraction family history and personal habits like smoking and alcohol were also studied. Coronary angiogram was done with multiple angulated views in Phillips Allura XD Single plane catheterisation lab.

Inclusion criteria

All patients with coronary artery disease who have undergone coronary angiogram in the prescribed time period, of all age groups and both sexes were included in the study.

Exclusion criteria

Procedures involving stand alone ventriclogram, aortogram, pacemaker implantation were excluded. Those who have undergone femoral access coronary angiogram. Those who have not given consent, Those who have undergone diagnostic angiogram and angioplasty simultaneously.

Results

Descriptive variables in baseline characteristics were measured as mean. Categorical valuables were measured as frequency. Our study population has 23 female patients and 54 male patients (Figure 2). That constitutes 29.9 % for female and 70.1 % for male patients. Personal habits of study population like alcoholics 19.5 %, smoking 26.0% smoking and alcoholic 19.5 and no bad habits 35.1 % (Figure 3). Co- morbid conditions Diabetes alone 22.1 %, Diabetes and systemic hypertension put together constitutes 22.1%, Systemic hypertension alone constitutes 27.3 %. Nil constitutes about 28.6%.(Figure 4) Altogether 77.9 % of patient has no family history of coronary artery disease. Rest of 22.1% patients has family history of coronary artery disease. Overall disease spectrum in our study Antero septal myocardial infarction 3.9 %

Figure 1

A: LAO Caudal view (Radial approach); B: LAO caudal view (Femoral approach)

https://s3-us-west-2.amazonaws.com/typeset-prod-media-server/78401be7-4e68-4b6c-87d9-15e61dfcfe9aimage1.png
Graph 1

Gender pie chart in numbers

https://s3-us-west-2.amazonaws.com/typeset-prod-media-server/78401be7-4e68-4b6c-87d9-15e61dfcfe9aimage2.png

Graph 2

Habbits in percentage

https://s3-us-west-2.amazonaws.com/typeset-prod-media-server/78401be7-4e68-4b6c-87d9-15e61dfcfe9aimage3.png

Graph 3

Co- morbidity prevalence in percentage

https://s3-us-west-2.amazonaws.com/typeset-prod-media-server/78401be7-4e68-4b6c-87d9-15e61dfcfe9aimage4.png

Graph 4

BMI division percentage

https://s3-us-west-2.amazonaws.com/typeset-prod-media-server/78401be7-4e68-4b6c-87d9-15e61dfcfe9aimage5.png

Table 1

Radiation exposure

Statistics

CUM AK

CUM DAP

FLOUR TIM

N

Valid

77

77

77

Missing

0

0

0

Mean

718.58247

43.55844

5.33468

Median

553.20000

30.70000

3.20000

Mode

304.500a

17.800a

1.400a

Std. Deviation

539.010607

36.350535

5.744874

Range

4041.500

223.000

39.010

Minimum

3.000

5.000

0.490

Maximum

4044.500

228.000

39.500

a. Multiple modes exist. The smallest value is shown

[i] CUM AK- Cumulative air kerma, CUM DAP- Cumulatice Dose Area product

[ii] FLOUR TIM- Flouro time.

Table 2

Cumulativedose area product and bod mass index crosstab

BMI

Total

<= 25.00

25.01+

CUM DAP

<= 30.000

17

21

38

30.001+

17

22

39

Total

34

43

77

[i] Chi square = 0.010, P=0.919, Not significant. BMI- Bod Mass Index,

[ii] CUM DAP – Cumulative Dose Area Product

Table 3

Fluoroscopy time and body mass index. Crosstab Count

BMI

Total

<= 25.00

25.01+

Flouro time

<= 3.200

19

20

39

3.201+

15

23

38

Total

34

43

77

[i] Chi square = 0.667, P=0.414, Not significant.

Anterior wall myocardial infarction 48.1%, Inferior wall myocardial infarction 33.8%, Non ST elevation myocardial infarction 6.5% and Unstable angina 7.8 %. Patient with body mass Index was divided into four groups and their distribution BMI a. < 21.00 -9.1 %, b. 21.01- 25.00- 35.1% c. 25.01-29.00 – 48.10 % d. 29.0+ - 7.8 % (Figure 5) Only less number of patients were in our study population beyond BMI 29 and less than 21 . Minimum age 34 years and maximum 73yrs mean age 42.08 years with Standard deviation being 10.872.

Minimum height of patients 145 cm and maximum being 175 cm. Mean value 158.4 cm and the standard deviation is 63.81. Weight of the study population minimum value 46 kg, maximum value 80 Kg. Mean value is 63.81 Kg and the standard deviation is 7.907.

Cumulative air kerma and cumulative dose area product, and fluoroscopy time are divided into below median and above median and then associated with Body mass Index.(Table 1)

On assessing cumulative dose area product for convenience of analysis in view of less study subjects in group d and a. Study population BMI is divided into two major groups i:e Below 25 and above 25. Cumulative dose area product into less than 30 Gy-cm 2 more than 30 Gy-cm2 We got the value of chi square = 0.01. and P= 0.919 which is not significant. (Table 1) Based on this analysis radiation exposure of patient is not associated with Body mass index within the BMI range of our study population.

Compared with BMI < 25, a patient BMI≥ 40 was associated with 2.1 fold increase in patient radiation dose and 7 fold increase in physician radiation dose.6 Increase in radiation dose is attributable to the increase energy required to overcome tissue attenuation and facilitates sufficient number of photons reaching the image intensifier to generate adequate images.

As per study by Ran D Madder. MD, Stacie Van Oosterhout. Med, Abbey Mulder, BSN. RN. Jan 2019. Circulation. Cardiovascular Interventions, DAP Vs BMI Group with BMI < 25 has DAP 44.5 Gy-cm2, 25 -29.9 DAP is 56 Gy- cm2,30- 34.9 DAP is 69.1 Gy- cm2, 35-39.9 DAP is 77.8 Gy- cm2, ≥ 40 DAP is 91.8 Gy- cm2 P value derived was < 0.001. In our study centre population none of the individual had BMI more than 35. Based on this, the subset of patients coming to our centre from nearby semi urban and rural areas yet to have considerable number of morbid obesity or class II obese patients, since most of them are leading physically active life style, epidemic of pandemic yet have its impact in this geographical area.

Body mass index was divided above and below 25 for convenience of calculation. Chi square value is 0.667 and the P= Value is 0.414. (Table 3) After confounding factors like catheter engagement difficult cases, radial spasm, anxious patients BMI was not correlating with fluoroscopy time in our study group. Our study group has predominantly between BMI of 21 and 29. A group and d group in the extreme of sample has less number. In study conducted by Ran D Madder as per reference 1, Fluoroscopy time in minutes with BMI value < 25.0 is 5.9 min, 25-30 is 5.6 min , 30-35 is 5.9 min, 35- 39.9 is 6.1 min, ≥ 40 is 7.0 with P Value of 0.26 1 which is not significant and correlates with our study.

Discussion

Long term radiation exposure among interventional cardiologists has been linked to multiple adverse effects7, 8 and the prevalence of obesity in cardiac catheterisation laboratory has been increased over time.2 Obesity is associated with increased health problems in patients. 10 Patient radiation exposure increases with body mass index especially with more than BMI 35.Body mass index is increasing the risk of physician radiation exposure as mentioned above and low sustained radiation exposure has inherent radiation hazards not only for the patients but also to the treating interventional cardiologist like cataract, carcinoma etc. Apart from radiation exposure BMI also has procedure related issues right from vascular access to outcome of procedure. Fluoroscopy time also increases, though statistically not significant proportionate to body mass index.

Conclusion

From our study patient radiation exposure as measured by Dose Area Product is incremental but statistically not significant enough to conclude positive correlation. As study is carried out in Tier II city, single centre study involving sample population who carry out physically active life style incidentally found to be predominantly between BMI of 21 to 29. Epidemic of obesity still to conquer this geographical area significant enough to produce coronary artery disease. Fluoroscopy time and cumulative air kerma also not significantly related to body mass index in the given range of our study

Limitations

Extremes of BMI group study sample is less to throw light on the effect of patient radiation exposure on body weight. Since it is a single centre study in Tier II city with small sample size, it is difficult to extrapolate the results to general population.

Conflict of Interest

None.

Source of Funding

None.

Acknowledgement

Dr. Suresh kumar. MD, DM. Assistant Professor of Cardiology. Chengalpattu Medical College. Chennai.

References

1 

A Shah P Das E Subkovas AN Buch M Rees C Bellamy Radiation dose during coronary angiogram: relation to body mass indexHeart Lung Circ2015241215

2 

ME Buscher D Smith D Share W Campbell S Mattichak M Sharma The burgeoning epidemic of morbid obesity in patients undergoing percutaneous coronary intervention: Insight from the Blue Cross Blushield of Michigen Cardiovascular ConsortiumJ Am Coll Cardiol201362868591

3 

JC Yanch RH Behrman MJ Hendricks JH Mccall Increased radiation dose to over weight and obese patients from radiographic examinationsRadiology2009252112839

4 

J Ector O Dragusin B Adriaenssens W Hubrechts R Willems H Ector Obesity is a major determinant of radiation dose in patients undergoing pulmonary vein isolation for atrial fibrillationJ Am Coll Cardiol200750323442

5 

A Shah P Das E Subkovas AN Buch M Rees C Bellamy Radiation dose during coronary angiogram: relation to body mass indexHeart Lung Circ2015241215

6 

RD Madder S VanOosterhout A Mulder TT Brock AT Clarey JL Parker Is the obesity epidemic impacting the occupational risk of physicians in the catheterisation laboratoryCirc Cardiovasc Interv2019121e00682310.1161/CIRCINTERVENTIONS.118.006823

7 

O Ciraj- Bjelac MM Rehani KH Sim HB Liew E Vano NJ Kleiman Risk for Radiation-Induced Cataract for Staff in Interventional Cardiology: Is There Reason for Concern?Catheter Cardiovasc Interv201076682634

8 

E Vano NJ Kleiman A Duran M M Rehani D Echeverri M Caberera Radiation cataract risk in interventional cardiology personnelRadiat Res201017444905

9 

JL Georges B Livarek H Gibault-Genty JP Messaoudi JL Aziza H Hautecoeur Variations of radiation dosage delivered to patients undergoing interventional cardiological procedures. A monocentric study 2002-05Arch Mal Coeur Vaiss2007100317581

10 

CJ Lavie PA Mcauley TS Church RV Milani SN Blair Obesity and cardiovascular diseases: implications regarding fitness, fatness, and severity in the obesity paradoxJ Am Coll Cardiol20146314134554

11 

S Balter Methods for measuring fluoroscopic skin dosePediatr Radiol200636Suppl 213640

12 

AJ Einstein KW Moser RC Thomson MD Cerqueria MJ Henzslova Radiation dose to patients from cardiac diagnostic imagingCirculation2007116111290305

13 

F Philiipe F Larrazet T Meziane A Dibie Comparison of transradial vs. transfemoral approach in the treatment of acute myocardial infarction with primary angioplasty and abciximabCatheter Cardiovas Interve20046116773

14 

M Sandborg SG Fransson H Petterson Evaluation of patient-absorbed doses during coronary angiography and intervention by femoral and radial artery accessEur Radiol20041446538

15 

G Christopoulos GE Christopoulos BV Rangan R Lane R Grabarkewitz D Haagan Comparison of radiation dose between different fluoroscopy systems in modern catheterisation laboratory. Results from bench test testing using an anthropometric phantomCatheter Cardiovasc Interv201586592732

16 

M Kastrati L Langen Brink M Piatkowski J Michaelsen D Reimann R Hoffmann Reducing radiation dose in coronary angiography and angioplasty using image noise reduction technologyAm J Cardiol201611833536

17 

AW Wassef B Hiebert A Ravandi J Ducas K Minhas M Vo Radiation dose reduction in the cardiac Catheterisation Laboratory utilizing a novel protocolJACC Cardiovasc Interv2014755507

18 

LR Dekker PH Van Der Voort TA Simmers XA Veerbeek RW Bullens MV Veer New image processing and noise reduction technology allows reduction of radiation exposure in complex electrophysiologic interventions while maintaining optimal image quality: a randomized clinical trialHeart Rhythm20131011167882

19 

T El-Sayed AS Patel JS Cho JA Kelly FE Ludwinski P Saha Radiation-Induced DNA Damage in Operators Performing Endovascular Aortic RepairCirculation201713625240616

20 

MG Andreassi E Piccaluga L Gargani L Sabatino A Borghini F Faita Subclinical carotid atherosclerosis and early vascular aging from long term low- dose ionizing radiation exposure: a genetic, telomere and vascular ultrasound study in cardiac catheterisation laboratory staffJACC Cardiovasc Interv20158461627

21 

MG Andreassi E Piccaluga G Guagliumi MD Greco F Gaita E Picano Occupational health risks in cardiac catheterisation laboratory workersCirc Cardiovasc Interv201694e00327310.1161/CIRCINTERVENTIONS.115.003273



jats-html.xsl

© 2024 Published by Innovative Publication Creative Commons Attribution 4.0 International License (creativecommons.org)